Tensile Slrcnglh, Tensile Straw, and Ullimalc Elongation
Low Tcmgcralurc Flex y_
Ham Shock
Coggcr Stability
Corrosive Effect
SPECIFICATION RT-1116 Issue 6 Page 3
If this document is printed it becomes uncontrolled. Check for the latest revision.
2005-2014 Tyco Electronics Corporation. All rights reserved
4.3.2 Tensile Strength, Tensile Stress, and Ultimate Elongation
Three specimens of tubing shall be tested for tensile strength, tensile stress, and ultimate elongation in
accordance with ASTM D 2671. For tubing size 3/8, the specimens shall be full sections of tubing. For sizes
1/2 and larger, the specimens shall be cut with Die D of ASTM D 412. The specimens shall have 1-inch (25-
mm) bench marks, centrally located. The testing machine shall have an initial jaw separation of 1 inch (25
mm) for full sections of tubing and 2 inches (50 mm) for die cut specimens. The rate of jaw separation shall
be 20 ± 2 inches (500 ± 50 mm) per minute.
4.3.3 Low Temperature Flexibility
Three specimens, each 12 inches (300 mm) in length, and a mandrel selected in accordance with Table 2,
shall be conditioned at -55 ± 3°C (-67 ± 5°F) for 4 hours. For tubing sizes 3/4 or less, the specimens shall be
whole sections of tubing recovered on a stranded wire (nearest AWG which is larger than the sleeving
maximum inside diameter after unrestricted shrinkage). For tubing sizes larger than 3/4, the specimens shall
be 1/4-inch (6.3-mm) wide strips cut from tubing which has been recovered in accordance with 4.3. After 4
hours conditioning, and while still at the conditioning temperature, the specimens shall be wrapped around
the mandrel for not less than 360° in 10± 2 seconds. The specimens then shall be visually examined for
evidence of cracking.
4.3.4 Heat Shock
Three 6-inch (150-mm) specimens of tubing shall be conditioned for 4 hours in a 225 ± 3°C
(437 ± 5°F) oven. After conditioning, the specimens shall be removed from the oven, cooled to room
temperature, and visually examined for evidence of dripping, flowing, or cracking.
4.3.5 Heat Resistance
Three specimens of tubing prepared and measured in accordance with 4.3.2, shall be conditioned for 168
hours in a 150 ± 2°C (302 ± 4°F) oven. After conditioning, the specimens shall be removed from the oven,
cooled to room temperature, and tested for tensile strength and ultimate elongation in accordance with 4.3.2.
4.3.6 Copper Stability
Three 6-inch (150-mm) specimens of tubing shall be slid over snug- fitting, straight, clean, bare copper
mandrels, either solid or tubular. The specimens on the mandrels shall be conditioned for 24 hours in a
desiccator or similar humidity chamber at 90 to 95 percent relative humidity and
23 ± 3°C (73 ± 5°F). The specimens on the mandrels then shall be conditioned for 168 hours in a 1
50 ± 2°C (302 ± 4°F) oven. After conditioning, the specimens shall be removed from the oven and cooled to
23 ± 3°C (73 ± 5°F). The copper mandrels shall then be removed from the tubing and the tubing and copper
mandrels shall be examined. Darkening of the copper due to normal air oxidation shall not be cause for
rejection. The tubing shall be tested for ultimate elongation in accordance with 4.3.2.
4.3.7 Corrosive Effect
The tubing shall be tested for copper mirror corrosion in accordance with ASTM D 2671, Procedure A. The
specimens shall be conditioned for 16 hours at 135 ± 3°C (275 ± 5°F).
4.3.8 Fluid Resistance
Six specimens, three 6-inch (150-mm) tubing specimens, which shall be weighed prior to immersion, and
three tensile specimens prepared and measured in accordance with 4.3.2, shall be immersed for 24 hours in
each of the test fluids listed in Table 3 at the temperature specified. The volume of the fluid shall not be less
than 20 times that of the specimens. After conditioning, all the specimens shall be lightly wiped and air dried
for 30 to 60 minutes at 23 ± 3°C (73 ± °F). The three specimens intended for the tensile strength and
elongation tests shall then be tested in accordance with 4.3.2. The other three specimens shall be weighed
before and after immersion and the weight change calculated as a percentage.
4.4 REJECTION AND RETEST
Failure of any sample of tubing to comply with any one of the requirements of this specification shall be
cause for rejection of the lot represented. Tubing which has been rejected may be replaced or reworked to
correct the defect and then submitted for acceptance. Before resubmitting, full particulars concerning the
rejection and the action taken to correct the defect shall be furnished to the inspector.